Notice of TR7A Series Warning Malfunction
The following problem has been found in the TR71A, TR72A and TR72A-S products.
Explanation of Problem
If you enable warning settings for both channel 1 and 2, once the first warning occurs on channel 1, warning monitoring will stop and warnings will not be issued after that on channel 1.
* Please note that warning monitoring and transmission for channel 2 will function normally even in this state.
* When warning settings are enabled for only one channel (Ch1 or Ch2), this problem will not occur.
Affected Products
TR71A/72A/72A-S : Firmware Versions 1.11 or lower
This issue will be resolved in the upcoming firmware update (Ver. 1.13), but please be aware of this issue until then. We apologize for any inconvenience caused by this problem.