RTR500BW/500BM/500BC

RTR500BW

RTR500BW is a Base Unit equipped with wired and wireless LAN functionality. Measurement data gathered via wireless communication from target Remote Units can be automatically uploaded to our cloud storage service T&D WebStorage Service. Remote monitoring, warning monitoring and device settings can also be carried out via the cloud service.
Also equipped with Bluetooth® and USB functions, it can be set on either a smartphone or a PC.

 Note

Here is a description of the device. For details about settings see [T&D 500B Utility Mobile App] or [RTR500BW for Windows].

Compatible Devices
Remote Units RTR500B Series Data Loggers:
 RTR501B/502B/503B/505B/507B (Including L Type)
RTR-500 Series Data Loggers:*
 RTR-574/576 (Including S Type),
 Discontinued: RTR-501/502/503/507S/505-Pt/505-TC/505-mA/505-V/505-P (Including L Type)
RTR-600 Series Data Loggers:*
 RTR-602S/602L/602ES/602EL
 Discontinued: RTR-601-110/130/E10/E30
Repeaters RTR500BC (Discontinued*: RTR-500)
* RTR-500 Series loggers and Repeaters, as well as, RTR-600 Series loggers do not have Bluetooth capability.
RTR500BW Part Names

Included: USB Cable US-15C, AC Adaptor AD-05A4 or AD-05C1, Registration Code Label, Manual Set (Warranty Included)

About the LED Display

LED(color) / Status Device Status
Power
(green)
ON Power is "ON"
Bluetooth Communication (blue) ON Bluetooth Communication is set to ON
BLINKING Bluetooth Communication in progress...
ACTIVE
(green)
ON Network communication available / Connected via USB
BLINKING Network or USB Communication in progress...
DIAG
(orange)
ON
  • Initializing after power on
  • Network transmission failure
W-LAN
(green)
ON Wireless LAN Communication Possible (Wired LAN Communication not available* )
BLINKING Wireless LAN Communication in progress...
ALARM
(red)
BLINKING Warning Issued (One of the following warnings was issued: Upper or Lower Limit Exceeded, Wireless Communication Error, Sensor Error, Low Battery)
Ethernet Port POE (orange) ON Power being supplied
Ethernet Port LINK (green) BLINKING LAN Communication in progress...
* Use by selecting either wired or wireless LAN. If wireless LAN is active, wired LAN cannot be used.
ACTIVE and DIAG Status Device Status
"ACTIVE" OFF, "DIAG" blinking
  • Time acquisition failure or time has not been set
  • No Remote Units have been registered
  • No settings for autonomic operations such as warning monitoring and sending current readings have been made
  • Other settings may be incomplete
"ACTIVE" and "DIAG" simultaneously blinking
  • Unable to connect to the wireless LAN access point
  • IP address cannot be received from DHCP server

External Output Terminal Specifications

By connecting a buzzer, siren, lamp, or other electromagnetic contactor to the external contact output terminal, it is possible to be notified when a warning occurs.

Use a flathead screwdriver to press down on the terminal button and attach the wire into the hole or detach it.
When attaching, pull the screwdriver away and make sure that the wire cannot become dislodged.
Voltage when OFF: AC/DC50V or less
Current when ON: 0.1A or less
Resistance when ON: 35Ω

Fit Wires Single wire Φ 1.0 (AWG18)
Twisted wire 0.75 ㎜2
Useable Wires Single wire Φ 0.4 - 1.0 ㎜ (AWG26 - 18)
Twisted wire 0.3 - 0.75 ㎜2 (AWG22 - 20)
Cable core diameter Φ 0.18 ㎜ or more
Covering removed about 10 ㎜

Power

When in use connect to the power supply using the supplied AC adaptor AD-05A4 or AD-05C1.
When using with a wired LAN, you can also supply power via LAN cable and a PoE (IEEE 802.3af) compatible hub.

  • Settings are retained even during a power failure and operation will resume when power is returned.

RTR500BW System Integration

T&D can provide interested users with RTR500BW communication specifications and file formats info for recorded data (TRZ or PRZ format) and current readings data. This information can be integrated into a customer developed system or application.
Please contact your local Distributor.

RTR500BM

The RTR500BM transmits the measured data collected wirelessly from target Remote Units to the cloud service T&D WebStorage Service via 4G data communication, which enables remote monitoring, warning monitoring, and device settings on the cloud.
It is also equipped with Bluetooth® and USB communication for making settings, which can be done via smartphone or PC.

 Note

Here is a description of the device. For details about settings see [T&D 500B Utility Mobile App] or [RTR500BM for Windows].

Compatible Devices
Remote Units RTR500B Series Data Loggers:
 RTR501B/502B/503B/505B/507B (Including L Type)
RTR-500 Series Data Loggers:*
 RTR-574/576 (Including S Type),
 Discontinued: RTR-501/502/503/507S/505-Pt/505-TC/505-mA/505-V/505-P (Including L Type)
Repeaters RTR500BC (Discontinued*: RTR-500)
* RTR-500 Series loggers and Repeaters do not have Bluetooth capability.
RTR500BM Part Names

Included: LTE Antenna CSR-0011, USB Cable US-15C, AA Alkaline Battery LR6(x4), AC Adaptor AD-05A3 or AD-05C1, Registration Code Label, Manual Set (Warranty Included)

SIM Card Preparation

Please purchase a SIM card that meets the following requirements:

  • Size: Nano SIM
    For the supported SIM cards, contact your local T&D distributor.
  • The card has been or can be activated
  • SIM which supports 4G data communication
    If you wish to receive warning messages via SMS, then you will need a card/contract supporting SMS.
    The SMS warning function is supported by firmware version R.01.00.06 or later
  • Contracted SIM card with a minimum speed of 200Kbps

When purchasing a SIM card, you may be asked for the IMEI number. The IMEI number for your RTR500BM can be found in the software/app as shown below:
T&D 500B Utility - [System] - [Device Info]
RTR500BM for Windows - [RTR500BM Settings Utility] – [Settings table] – [Version]

SIM Card Installation/Removal

 IMPORTANT
  • When inserting or removing a SIM card, avoid exposure to static electricity.
  • Before inserting or removing the card, make sure to disconnect all cables (AC adaptor, USB cable, and external power cable).
  • Be careful not to touch or scratch the IC area of the card. This may cause poor contact.
  • Make sure to insert the card in the proper direction.

Remove the battery cover to find the SIM card holder.
Place the SIM card with the IC (electrode) side down and the notch to the lower left, and slide it to the left.

  • Slide it completely left until you hear a "click" sound. Incomplete or improper insertion may cause insufficient contact with the IC and result in a communication error.
  • To remove the SIM card, press in with your fingernail to release the lock, then carefully slide the card out of the SIM card holder.

Power

When in use connect to the power supply using the supplied AC adaptor AD-05A3 or AD-05C1.
The included batteries will serve as a backup power source in the event of a power failure. If running only on batteries, the estimated battery life is about 2 days under the following conditions: only one Remote Unit and no Repeaters, downloading data once a day, sending current readings at 10-min interval.
The optional battery connection adaptor BC-0204 can be used to connect to an external power source such as a car battery.

Power Source Conditions (BC-0204)

External Battery Connection Adaptor (BC-0204)
Current:

MAX 2A

Voltage:

input DC9 – 38V
output DC5V

 IMPORTANT

Cautions for Changing the Batteries

  • If the PWR LED is blinking, a communication is in progress and the batteries should NOT be changed at this time.
  • Do not insert or change batteries with wet hands.
  • When inserting make sure no water or foreign objects get inside the case.
  • Make sure that + and − are in the correct direction.
  • Make sure to use new batteries of the same kind.
  • When using batteries as a backup power source in case of power failure, we recommend that you change the batteries every two or three years. Leaving batteries in a device for a long period of time may cause battery leakage and corrosion.

Connect the Antennas

Remove the protective covers from the antenna connectors (LOCAL and CELLULAR) on the unit, and connect the supplied LTE Antenna CSR-0011.

 Note

If you wish to use the GPS function, please purchase a GPS antenna in advance. (T&D Corporation does not handle or sell GPS antennas.) Remove the protective cover from the antenna connector (GPS) on the unit, and connect the GPS antenna.
GPS Antenna Connector: SMA Female Jack
Power Supply: 3.3V

LED Display and Operation Switch

Status Details
PWR
(Power)
Green
BLINKING Running on battery power only
ON Running on AC Adaptor or external power source
Connected via USB
BLINKING
(rapidly)
During communication via mobile network, short range radio communication, or USB connection
OFF In low energy consumption mode (functions not operable)
DIAG
(Diagnosis)
Orange
ON No SIM card inserted / Poor SIM card contact
BLINKING Starting up after power on
No Remote Units have been registered
Auto-download of recorded data cannot be carried out due to other improperly made settings or unmade settings
ALM
(Alarm)
Red
BLINKING A measurement has exceeded one of the set limits
  • The contact input is ON
  • There is trouble in the Remote Unit (low battery, poor sensor connection, etc.)
  • Wireless communication with the Repeater or Remote Unit failed
ANT. LEVEL
(4G Network Reception Level)
Green
ON 3: Strong
2: Average
1: Weak
OFF Outside of communication range
Bluetooth
Blue
ON Bluetooth function ON
OFF Bluetooth function OFF
BLINKING BLINKING: During Bluetooth communication with a mobile device
BLINKING (rapidly): During command communication execution
  • To ensure stable 4G (LTE) mobile signals, place the device in a location where at least two ANT. LEVEL LEDs light up.
  • Even if the LED display is normal, the connection to the access point may have failed, e.g.,, the SIM card has not been activated. Therefore, make sure to carry out transmission tests before the operation.
Operation Switch
Run The following functions are operable: Auto-Downloading and Sending of Recorded Data, Warning Monitoring, and Auto-Sending of Current Readings.
STBY (Standby) The unit is in a low energy consumption mode and functions are not operable.
In order to prevent unnecessary or unexpected data transmission, please leave the operation switch to STBY position until the units have all been set up and ready for communication See: [RTR500BM for Windows] - [Device Installation] section of this Help

External Input/Output Terminal

By connecting a buzzer, siren, lamp, or other electromagnetic contactor to the external contact input/output terminal, it is possible to be notified when a warning occurs.
The optional alarm connection cable (AC0101) is available. Please contact your local distributor for purchase.

Contact Input (for external alarm input) Input Terminal Internal Pull-up: 3 V 100 KΩ
Maximum Input Voltage: 30 V
GND  
Contact Output (OUT) Output Terminal (for external alarm output) PhotoMOS Relay Output
OFF-State Voltage: AC/DC 50 V or less
ON-State Current: 0.1 A or less
ON-State Resistance: 35 Ω
GND  

RTR500BM System Integration

T&D can provide interested users with RTR500BM communication specifications and file formats info for recorded data (TRZ format) and current readings data. This information can be integrated into a customer developed system or application.
Please contact your local Distributor.

RTR500BC

Using as a Base Unit

Connect the RTR500BC to your computer with the included USB cable and collect data via wireless communication.

 Note

Here is a description of the device. For details about settings see [RTR500BC for Windows].

Compatible Devices
Remote Units RTR500B Series Data Loggers:
 RTR501B/502B/503B/505B/507B (Including L Type)
RTR-500 Series Data Loggers:
 RTR-574/576 (Including S Type),
 Discontinued: RTR-501/502/503/507S/505-Pt/505-TC/505-mA/505-V/505-P (Including L Type)
RTR-600 Series Data Loggers*:
 RTR-602S/602L/602ES/602EL,
 Discontinued: RTR-601-110/130/E10/E30
Repeaters RTR500BC, (Discontinued: RTR-500)
* Customers wishing to use the RTR500BC as a Base Unit in conjunction with the RTR-600 series devices, please contact your local distributor for the communications protocol specifications to write your own software.

Using as a Repeater

If short range wireless communication cannot successfully be carried out due to obstacles, or if you wish to extend the range, Repeater(s) can be added between Base and Remote Unit(s).

  • To register a device as a Repeater use the dedicated application for the Base Unit being used.
Compatible Devices
Base Units RTR500B Series: RTR500BW, RTR500BM, RTR500BC
RTR-500 Series (Discontinued): RTR-500NW, RTR-500AW, RTR-500MBS-A, RTR-500, RTR-500DC
RTR500BC Part Names

Included: USB Cable US-15C, Manual Set (Warranty Included)

Power

When using the RTR500BC as a Repeater, the power supply options are as follows.

  • Alkaline battery (not included): It can be run on 2 AA alkaline batteries for about 6 months (when wireless communication is carried out for 5 minutes every day). Battery life shortens depending on the frequency of communication.
  • Optional AC adaptor AD-06A1 or AD-06C1
  • Optional battery connection adaptor BC-0204: Can be used to connect to an external power source such as a car battery.

When using the RTR500BC as a Base Unit, it works on the USB bus power and it is not necessary to use another power source.

Power Source Conditions (BC-0204)

External Battery Connection Adaptor (BC-0204)
Current:

MAX 2 A

Voltage:

input DC9 - 38 V
output DC5 V

 IMPORTANT

Cautions for Changing the Batteries

  • If the LED is blinking a communication is in progress and the batteries should NOT be changed at this time.
  • Do not insert or change batteries with wet hands.
  • When inserting make sure no water or foreign objects get inside the case.
  • Please make sure that +/− are in the correct directions.
  • Make sure to use new batteries of the same kind that are within expiration date.
  • When using batteries as a backup power source in case of power failure, we recommend that you change the batteries every two or three years. Leaving batteries in a device for a long period of time may cause battery leakage and corrosion.

About the LED Display

BLINKING (Green)

Wireless Communication in progress...

BLINKING (Blue)

Bluetooth Communication in progress...

LCD Display

Communication Status

Will blink when either Wireless Communication or Bluetooth Communication is in progress.

Passcode Lock

Will appear when the Bluetooth Passcode is set.

Power Supply Status

Will appear while connected to an external power source, such as USB, AC adaptor or car battery.

Battery Warning

BLINKING: Will blink when battery power is becoming low.
Lamp ON: This appears when the battery level is too low to operate. Please change the batteries immediately.
(See: [Power]-IMPORTANT (Cautions for Changing the Batteries)

Device Info / Operational Messages

Here is displayed a character string that identifies the device as a Base Unit or a Repeater, and also operational messages.
For details see [Operational Messages] below.

Operational Messages

Base Unit

This will appear when the device has been registered as a Base Unit or prior to being registered as a Repeater.

Repeater

This will appear after the device has been registered as a Repeater.
[RPT]-[GR]-[No] displayed in a loop.

  • [GR] is a user selected alphanumeric name that can only be set when the RTR500BW or RTR500BM is registered as a Base Unit. It can be used to identify the device or route and so on. If no settings have been made [RPT]-[No] will be alternatingly displayed.
  • [No] is a number automatically assigned when a Repeater is registered in the order of registration.

Unbootable Status

This will appear when there is not enough battery voltage to boot the device.

Device Start

This will be displayed under the following conditions:

  • When power is turned on from a state where there was no power.
  • When a Reset Command was sent by the application
  • When a firmware update has been completed

Firmware Update in Progress

When a firmware update is in progress the following will alternatingly displayed:

Checking the Battery Level

Battery level can be checked via Settings Utility software for the Base Unit.

It can also be checked on a smartphone via Bluetooth

You can check the battery level of a device in the Settings Menu of the T&D 500B Utility by going to [Base Unit / Repeater Settings] - [Wireless Route] tab.

Check by connecting to a PC via USB

Connect the Repeater directly to a PC with a USB cable and in the Settings Utility for the Base Unit you are using you can find info for registered Repeaters. The battery level will appear as a number from 0-5.

RTR500BC System Integration

T&D can provide interested users RTR500BC communication specifications and file formats info for recorded data (TRZ format) and current readings data. This information can be integrated into a customer developed system or application.
Please contact your local Distributor.