Support Information
Update Report
March 11, 2011
TR-57DCi Firmware Update Info (Ver. 1.3 to 1.4)
In order to fix the following problems in the TR-57DCi, we have released this update program for the firmware.
Affected Products and Conditions
Device:
TR-57DCi
Firmware:
TR-57DCi Ver. 1.3
Modification Details
From TR-57DCi Firmware (Ver. 1.3) to (Ver. 1.4)
- An error was corrected that occurred when downloading data via optical communication, whereby if you pressed the <ON/OFF> button to turn off the TR-57DCi, optical communication could not be carried out even after having turned on the power.
- An error was corrected that occurred when the recorded data stored in the TR-57DCi exceeded 128 data sets and data stored after that could not be selected correctly for graph display.
- Devices were added for which the user can specify period of data to be downloaded.
(TR-73U has not been added yet.)