| Measurement Items |
Temperature |
Humidity |
| Measurement Range |
-30 to 80°C |
0 to 99%RH |
| Measurement Resolution |
0.1°C |
0.1 %RH |
| Measurement Accuracy |
±0.3°C (at 10 to 40 °C)
±0.5°C (at all other temperatures) |
±2.5 %RH(at 25 °C, 10 to 85 %RH)
±4.0 %RH(at 25 °C, 0 to 10 % or 85 to 99 %RH)
At temperatures other than 25 °C and ≥ 0 °C, add ±0.1 %RH per degree of difference from 25. |
| Response Time (90%) |
Approx. 7 min. |
Approx. 20 sec. |
| Sensor Temperature Durability |
-30 to 80°C |
| Long Term Stability |
±1%RH/yr, ±0.1°C/yr |
| Humidity Hysteresis |
- |
1.5%RH or lower (*1) |
| Sensor Dimensions |
H 19 mm x W 65 mm x D 10 mm |
| Cable Length |
1.5 meters |
| Conditions for Use |
Do not expose to condensation, dampness, corrosive gases or organic solvents |
|
*1: When used at temperatures equal to or higher than 0°C. |
Materials: ①Polycarbonate ②Vinyl chloride-shielded wire