New Releases

New Product

Announcing Additional Models to Our TR-5i Series - The "Data Recorder TR-55i" Lineup

Date Issued: April 13, 2012

The TR-55i models are designed to measure and record temperature (Thermocouple, PT100/Pt1000), voltage, 4-20mA, or pulse count. They can measure and record temperature in a wide range using a thermocouple or Pt100/Pt1000 sensor, or other signal types such as 4-20mA and 0-5V. The body of the data loggers are water resistant and can be used in a environment of -40°C to 80°C. Moreover, battery-powered operation means the loggers can be placed anywhere without the need for AC power. The data recorded by the TR-55i loggers can be transfered to PC whereby users can analyze the data and view in graph form.

Features

Compatible Data Collection Devices

Communication Port: TR-50U2, TR-50U
Data Collector: TR-57DCi

Compatible software for TR-55i models is “T&D Recorder for Windows (TR-5, 7xU)” version 2.00 or later.

TR-55i Lineup

news image

Application Examples

news image

Specifications

  TR-55i-TC TR-55i-Pt TR-55i-V TR-55i-mA TR-55i-P
Measurement
Items
Temperature 1ch Temperature 1ch Voltage 1ch 4-20mA 1ch Pulse Count 1ch
Sensor Thermocouple: Type K, J, T, S Pt100, Pt1000
(3-wire)
- - -
Measurement
Range
-199 to 1700 °C -199 to 600 °C 0 to 22 V 0 to 20 mA
(Operational up to 40 mA)
Input Signal:
・Non-voltage Contact Input
・Voltage Input (0 to 27 V)

Detection Voltage
Lo: 0.5 V or less
Hi: 2.5V or more

Chattering Filter:
ON: 15 Hz or less
OFF: 3.5 kHz or less
Accuracy Thermocouple Measurement
±0.3 °C + 0.3 % of reading (Type K, J, T)
±1 °C + 0.3 % of reading (Type S)

Cold Junction Compensation
±0.3 °C (10 to 40°C)
±0.5 °C (-40 to 10 °C / 40 to 80 °C)
±0.3°C + 0.3 % of reading (10 to 40 °C)
±0.5°C + 0.3 % of reading (-40 to 10 °C / 40 to 80 °C)
±0.5 mV + 0.3 % of reading (10 to 40 °C)
±1 mV + 0.5 % of reading (-40 to 10 °C / 40 to 80 °C)
±0.05 mA + 0.3 % of reading (10 to 40 °C)
±0.1mA + 0.3 % of reading (-40 to 10 °C / 40 to 80 °C)
Note: The above mentioned temperature is the operating environment for Input Module only.
Logging
Capacity
16,000 readings
Recording
Interval
Select from 15 choices: 1, 2, 5, 10, 15, 20, 30 sec. or 1, 2, 5, 10, 15, 20, 30, 60 min.
Power Lithium Battery: LS14250 or CR2 x 1
Battery Life*1 Approx. 14 months (10 months with Infrared Comm. Enabled) Approx. 24 months (14 months with Infrared Comm. Enabled) Approx. 16 months (11 months with Infrared Comm. Enabled) Approx. 16 months (11 months with Infrared Comm. Enabled) Approx. 16 to 24 months (11 to 18 months with Infrared Comm. Enabled)
Operating
Environment
-40 to 80 °C
Waterproof
Capacity
IP64: Splash proof (rated for use in daily life) Note: Input Module is not water resistant.
Data Collection
Devices
Communication Port: TR-50U2, TR-50U
Data Collector: TR-57DCi

*1: Battery life varies depending upon the ambient temperature in which it is used, the recording interval, the frequency of communication, and the battery performance. All estimates are based on operations carried out with a new battery and are in no way a guarantee of actual battery life. When infrared communication function is enabled, battery life may be shortened if the unit is used under the inverter type fluorescent lighting.

Release Date

Apr. 2012